EPA-OW: 218.7:  Hexavalent Chromium in Water by Ion Chromatography

  • Summary
  • Analytes
  • Revision
  • Data and Sites
Official Method Name
Method 218.7: Determination of Hexavalent Chromium in Drinking Water by Ion Chromatography with Post-Column Derivatization and UV–Visible Spectroscopic Detection
Current Revision
Version 1.0, November 2011
Media
WATER
Instrumentation
Ion Chromatography with Post-Column Derivatization and UV–Visible Spectroscopic Detection
Method Subcategory
Inorganic
Method Source
  EPA-OW
Citation
  S. Wendelken, G. Smith, D. Munch, A. Zaffiro and M. Zimmerman, 2011, Method 218.7: Determination of Hexavalent Chromium in Drinking Water by Ion Chromatography with Post-Column Derivatization and UV–Visible Spectroscopic Detection, U.S. EPA Office of Water: Version 1.0 November 2011, 31 p.
Brief Method Summary
Samples are preserved with a combined buffer/dechlorinating reagent which complexes free chlorine and increases the pH to a value greater than eight. A measured volume (usually 1 mL) of the sample is introduced into an ion chromatograph. CrO42- is separated from other matrix components on an anion exchange column. CrO42- is derivatized with 1,5-diphenylcarbazide in a post-column reactor and is detected spectrophotometrically at a wavelength of 530 nm. Cr(VI) is qualitatively identified via retention time, and the concentration of CrO42- in the sample is calculated using the integrated peak area and the external standard technique. Results are reported in units of ug/L of Cr(VI).
Scope and Application
Method 218.7 provides procedures for the determination of hexavalent chromium Cr(VI) as the chromate anion CrO42- in finished drinking water using ion chromatography. Samples are analyzed by direct injection. This method is intended for use by analysts skilled in the operation of ion chromatographic instrumentation and in the interpretation of the associated data.
Applicable Concentration Range
0.012 to 0.036
Interferences
Use high-density polyethylene (HDPE) or polypropylene copolymer sample bottles. Reagents and the ion chromatographic system must be routinely demonstrated to be free from interferences. To ensure sample integrity, Cr(VI) must be protected from reduction, and Cr(III), if present, must not oxidize to Cr(VI) during sample storage.
Quality Control Requirements
QC criteria discussed in the following sections are summarized in Section 17, Tables 8 and 9.
Sample Handling
Samples are stable for at least 14 days at both ambient temperature 25oC and chilled temperature 6oC.
Maximum Holding Time
14 days
Relative Cost
Unknown
Sample Preparation Methods
The samples are preserved with a combined buffer/dechlorinating reagent. Either the liquid formulation or the solid formulation of the preservative described in the following sections may be used. Open the tap and allow the system to flush for approximately 5 minutes. Fill sample bottles with 100-mL of sample, taking care not to flush out the preservative. Invert the bottle several times to mix the sample with the preservative.