Hach Co.: 8156: pH in Water
Official Method Name
|
pH in Water Using a Portable Hach One Meter; Two-Standard Calibration in Automatic Mode With Temperature Probe |
---|---|
Current Revision
| 1997 |
Media
|
WATER |
Instrumentation
|
pH Meter |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
|
Brief Method Summary
|
This method requires the use of an electrode and pH meter. |
Scope and Application
|
This method is applicable to determining pH in drinking water, surface and ground waters, and domestic and industrial wastewaters. |
Applicable Concentration Range
|
None given. |
Interferences
|
Sodium error, usually present in alkaline solutions, is low, but increases at pH values higher than pH 11. |
Quality Control Requirements
|
Not included. |
Sample Handling
|
Collect samples in plastic or glass bottles. Fill completely and cap tightly. Refrigerate at 4oC and analyze within 6 hours. |
Maximum Holding Time
|
6 hours |
Relative Cost
|
Less than $50 |