EPA-EAD: 604: Phenols in Water Using GCECD/FID
Official Method Name
|
Phenols |
---|---|
Current Revision
| 40 CFR Part 136, Appendix A (Current Edition) |
Media
|
WATER |
Instrumentation
|
Gas Chromatography with Electron Capture Detection |
Method Subcategory
|
Organic |
Method Source
|
|
Citation
|
|
Brief Method Summary
|
A 1-L sample is acidified and extracted with methylene chloride using a separatory funnel. The methylene chloride extract is dried and exchanged to 2-propanol during concentration to a volume of 10-mL or less. The concentrations of phenols in the extract are measured using a gas chromatography (GC) system equipped with a flame ionization detector (FID). The method also provides for a derivatization and column chromatography cleanup procedure to aid in the elimination of interferences. Using this approach, the concentrations of phenols in the extract are measured by detecting derivatives using a GC system with an Electron Capture Detector (ECD). |
Scope and Application
|
This method covers the determination of phenol and certain substituted phenols in wastewater and other waters. |
Applicable Concentration Range
|
Ranges differ for each analyte depending on matrix and interferences. |
Interferences
|
(A) Glassware contamination: Thoroughly clean glassware, including baking or solvent rinse. (B) Reagent contamination: Use high purity reagents. (C) Extracted interferences: Interference from extracted non-target compounds, with retention times similar to target compounds, can be reduced by performing the derivitization cleanup procedure in the method. (D) Basic-sample wash: The wash can significantly reduce recovery of phenol and 2,4-dimethylphenol. |
Quality Control Requirements
|
Each laboratory that uses this method is required to operate a formal quality control program. The minimum requirements of this program consist of an initial demonstration of laboratory capability and an ongoing analysis of spiked samples to evaluate and document data quality. The laboratory must maintain records to document the quality of data that is generated. Ongoing data quality checks are compared with established performance criteria to determine if the results of analyses meet the performance characteristics of the method. When results of sample spikes indicate atypical method performance, a quality control check standard must be analyzed to confirm that the measurements were performed in an in-control mode of operation. |
Sample Handling
|
Grab samples must be collected in glass containers. [Note: Do NOT pre-rinse bottles]. Composite samples should be collected in refrigerated glass containers in accordance with the requirements of the program. Automatic sampling equipment must be as free as possible of Tygon tubing and other potential sources of contamination. All samples must be iced or refrigerated at 40C from the time of collection until extraction. Fill the sample bottles and, if residual chlorine is present, add 80 mg of sodium thiosulfate per liter of sample and mix well. EPA Methods 330.4 and 330.5 may be used for measurement of residual chlorine. Field test kits are available for this purpose. |
Maximum Holding Time
|
Extract w/in 7 days; Analyze extracts w/in 40 days. |
Relative Cost
|
$201 to $400 |
Sample Preparation Methods
|