EPA-NERL: 160.4: Volatile Residue by Muffle Furnace
Official Method Name
|
Residue, Volatile (Gravimetric, Ignition at 550 oC) |
---|---|
Current Revision
| Issued 1971 |
Media
|
WATER |
Instrumentation
|
Gravimetry |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
|
Brief Method Summary
|
The residue obtained from the determination of total, filterable or non-filterable residue is ignited at 550oC in a muffle furnace. The loss of weight on ignition is reported as mg/L volatile residue. |
Scope and Application
|
This method determines the weight of solid material combustible at 550oC. The test is useful in obtaining a rough approximation of the amount of organic matter present in the solid fraction of sewage, activated sludge, industrial wastes, or bottom sediments. |
Applicable Concentration Range
|
Not Available. |
Interferences
|
None. |
Quality Control Requirements
|
None. |
Sample Handling
|
Preservation of the sample is not practical; analysis should begin as soon as possible. Refrigeration or icing to 4oC is recommended. |
Maximum Holding Time
|
7 days (MCAWW, Table 1). |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|
None. |