EPA-NERL: 273.1: Sodium by Flame AA
Official Method Name
|
Sodium (Atomic Absorption, Direct Aspiration) |
---|---|
Current Revision
| Issued 1971; Editorial Revision 1974 |
Media
|
WATER |
Instrumentation
|
Flame Atomic Absorption |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
|
Brief Method Summary
|
A sample prepared according to EPA Method 200.0 is analyzed using a flame atomic absorption spectrometer. Specifications for instrument operation and preparation of standards are provided in the method. |
Scope and Application
|
This method determines sodium by flame atomic absorption spectrometry. Sample pretreatment according to EPA Method 200.0 determines the applicability. |
Applicable Concentration Range
|
0.03 - 1.0 mg/L (optimal range); 0.015 mg/L (sensitivity). |
Interferences
|
For interferences see EPA Method 200.0. Low-temperature flames increase sensitivity by reducing ionization of this metal. Potassium may also be added to control ionization. |
Quality Control Requirements
|
QC requirements for drinking water procedures are provided in method 200.0 |
Sample Handling
|
Sample handling and preservation procedures are provided in EPA Method 200.0. |
Maximum Holding Time
|
6 months (MCAWW, Table 1). |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|
200.0 |