EPA-NERL: 283.1: Titanium by Flame AA
Official Method Name
|
Titanium (Atomic Absorption, Direct Aspiration) |
---|---|
Current Revision
| Issued 1974 |
Media
|
WATER |
Instrumentation
|
Flame Atomic Absorption |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
|
Brief Method Summary
|
A sample prepared according to EPA Method 200.0 is analyzed using a flame atomic absorption spectrometer. Specifications for instrument operation and preparation of standards are provided in the method. |
Scope and Application
|
This method determines titanium by flame atomic absorption spectrometry. Sample pretreatment according to EPA Method 200.0 determines the applicability. |
Applicable Concentration Range
|
5 - 100 mg/L (optimal range); 2.0 mg/L (sensitivity). |
Interferences
|
For interferences see EPA Method 200.0. A number of elements increase the sensitivity of titanium. To control this problem, potassium must be added to standards and samples alike. |
Quality Control Requirements
|
QC requirements for drinking water procedures are provided in method 200.0 |
Sample Handling
|
Sample handling and preservation procedures are provided in EPA Method 200.0. |
Maximum Holding Time
|
6 months (MCAWW, Table 1). |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|