USGS-NWQL: I-4602: Phosphorus, orthophosphate plus hydrolyzable, total, colorimetric, ASF
Official Method Name
|
Phosphorus, orthophosphate plus hydrolyzable, colorimetric,phosphomolybdate, automated-segmented flow |
---|---|
Current Revision
| 1973 |
Media
|
WATER |
Instrumentation
|
Automated Spectrophotometer |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
Methods for the Determination of Inorganic Substances in Water and Fluvial Sediments, Techniques of Water-Resources Investigations of the United States Geological Survey, Book 5, Chapter A1 Edited by Marvin J. Fishman and Linda C. Friedman |
Brief Method Summary
|
Polyphosphates (P2O7)-4, (P3O10)-5, etc., and a few organic phosphorus compounds are converted to orthophosphate by an acid hydrolysis. Orthophosphate ion reacts with ammonium molybdate in acidic solution to form phosphomolybdic acid, which, upon reduction with ascorbic acid, produces an intensely colored blue complex. Antimony potassium tartrate is added to increase the rate of reduction (Murphy and Riley, 1962; Gales and others, 1966). Mercuric chloride-preserved samples are fortified with a minimum of 85 mg/L NaCl to overcome the interference from mercury in the analysis. |
Scope and Application
|
This method may be used to analyze most water-suspended sediment containing between 0.01 and 1.0 mg/L combined acid hydrolyzable and orthophosphate phosphorus. Samples containing greater concentrations need to be diluted. |
Applicable Concentration Range
|
0.01 to 1.0 mg/L |
Interferences
|
Barium, lead, and silver interfere by forming a phosphate precipitate, but the effect is usually negligible in natural waters. The interference from silica, which forms a pale-blue complex, is slight and can be considered negligible. Nitrite interferes, but can be oxidized to nitrate with hydrogen peroxide before analysis. Residual chlorine must be removed by boiling the sample. |
Quality Control Requirements
|
Calibrate instrument using calibration standards (CAL); quality control samples (QCS); and laboratory blanks (LB) analyzed at a minimum of 1 for every 10 samples. |
Sample Handling
|
Container Description: 125 mL brown polyethylene bottle. Treatment and Handling: Use unfiltered sample to rinse container; preservation with mercuric chloride; chilled to 4 degrees C and ship immediately. |
Maximum Holding Time
|
30 days |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|