USGS-NWQL: I-5900: Zinc, recoverable-from-bottom-material, dry wt, atomic absorption spectrometric
Official Method Name
|
Zinc, atomic absorption spectrometric, direct |
---|---|
Current Revision
| 1972 |
Media
|
SOILS/SEDIMENT |
Instrumentation
|
Flame Atomic Absorption |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
Methods for the Determination of Inorganic Substances in Water and Fluvial Sediments, Techniques of Water-Resources Investigations of the United States Geological Survey, Book 5, Chapter A1 Edited by Marvin J. Fishman and Linda C. Friedman |
Brief Method Summary
|
Zinc is determined by atomic absorption spectrometry by direct aspiration of the sample into an air-acetylene flame. The procedure may be automated by the addition of a sampler and either a strip-chart recorder or a printer or both. |
Scope and Application
|
This method may be used to analyze bottom material containing at least 1.0 ug/g of zinc. Sample solutions containing more than 500 ug/L of zinc need to be diluted or less scale expansion used. Recoverable zinc in bottom material needs to undergo preliminary digestion-solubilization by method I-5485 before being determined. |
Applicable Concentration Range
|
10 to 500 ug/L |
Interferences
|
Magnesium at concentrations greater than 100 mg/L interferes unless other cations, such as sodium, are present in the sample. Individual concentrations of sodium, potassium, sulfate, chloride (9,000 mg/L of each), calcium (4,500 mg/L), nitrate (2,000 mg/L), iron (4 X 106 ug/L), and cadmium, nickel, copper, lead, cobalt, and chromium (10,000 ug/L each) do not interfere. Greater concentrations of each constituent were not investigated. Samples containing 100 mg/L of silica cause no interference; however, zinc recovery is approx. 10 percent low in samples containing 200 mg/L of silica. |
Quality Control Requirements
|
Calibrate instrument using calibration standards (CAL); quality control samples (QCS); and laboratory blanks (LB) analyzed at a minimum of 1 for every 10 samples. |
Sample Handling
|
Bottom Material (stream bed sediment) 500 mL wide mouth propylene bottle; no treatment or preservation requirements. |
Maximum Holding Time
|
|
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|
I-5485 |