USGS-NWQL: I-1054: Aluminum, dissolved, water, DCP
Official Method Name
|
Aluminum, atomic emission spectrometric, d-c plasma |
---|---|
Current Revision
| Revised 1985 |
Media
|
WATER |
Instrumentation
|
Automated Spectrophotometer |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
Methods for the Determination of Inorganic Substances in Water and Fluvial Sediments, Techniques of Water-Resources Investigations of the United States Geological Survey, Book 5, Chapter A1 Edited by Marvin J. Fishman and Linda C. Friedman |
Brief Method Summary
|
Aluminum is determined by a direct-reading emission spectrometer which utilizes a d-c argon plasma as an excitation source (Johnson and others, 1979 a,b, 1980). A mixture of lithium chloride, sulfuric acid, and glycerin is added to samples and standards to provide a common background matrix and to compensate for viscosity changes. The liquid mixture is then converted by a ceramic nebulizer into a fine aerosol and introduced into the plasma via a plastic spray chamber and Pyrex injection tube. Aluminum is determined on the basis of the average of two replicate exposures, each of which is performed on a 10-second integrated intensity. Calibration is performed by standardization with a high-standard solution and a blank. |
Scope and Application
|
This method may be used to analyze finished water, natural water, industrial water, and water-suspended sediment containing from 10 to 1000 ug/L of aluminum. Samples containing more than 1000 ug/L aluminum and (or) with specific conductances greater than 10,000 uS/cm need to be diluted. |
Applicable Concentration Range
|
10-1000 ug/L |
Interferences
|
Stray-light effects in a high-resolution, single-element d-c argon plasma emission spectrometer are negligible. |
Quality Control Requirements
|
Calibrate instrument using calibration standards (CAL). Quality control samples (QCS) and laboratory blanks (LB) analyzed at a minimum of 1 each after every 10 samples. (Reference OFR 95-443). |
Sample Handling
|
Container Description: 250 mL Polyethylene bottle, acid-rinsed. Treatment and Preservation: Filter through 0.45-um filter, use filtered sample to rinse containers and acidify sample with HNO3 to pH < 2. |
Maximum Holding Time
|
180 days |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|