USGS-NWQL: I-5135: Cadmium, recoverable-from-bottom-material, various, FLAA
Official Method Name
|
Cadmium, atomic absorption spectrometric, direct |
---|---|
Current Revision
| Revised 1985 |
Media
|
VARIOUS |
Instrumentation
|
Flame Atomic Absorption |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
Methods for the Determination of Inorganic Substances in Water and Fluvial Sediments, Techniques of Water-Resources Investigations of the United States Geological Survey, Book 5, Chapter A1 Edited by Marvin J. Fishman and Linda C. Friedman |
Brief Method Summary
|
Cadmium is determined by atomic absorption spectrometry by direct aspiration of the sample into an air-acetylene flame without preconcentration or pretreatment of the sample. |
Scope and Application
|
This method may be used to analyze water and water-suspended sediment containing at least 10 ug/L of cadmium. Sample solutions containing more than 250 ug/L need either to be diluted or to be read on a less expanded scale. Sample solutions containing less than 10 ug/L need to be analyzed by the atomic absorption spectrometric chelation-extraction method, providing that the interference limits discussed in that method are not exceeded. This method may be used to analyze bottom material containing at least 1 ug/g of cadmium. Total recoverable cadmium in water-suspended sediment needs to undergo preliminary digestion-solubilization by method I-3485, and recoverable cadmium in bottom material needs to undergo preliminary digestion-solubilization by method I-5485 before being determined. |
Applicable Concentration Range
|
> 1 ug/g |
Interferences
|
Individual concentrations of sodium (9,000 mg/L), potassium (9,000 mg/L), magnesium (4,500 mg/L), sulfate (9,000 mg/L), chloride (9,000 mg/L), nitrate (100 mg/L), iron (4 X 106 ug/L), and cobalt, nickel, copper, zinc, lead, and chromium (10,000 ug/L each) do not interfere. Greater concentrations of each constituent were not investigated. Calcium at concentrations greater than 1,000 mg/L suppresses the cadmium absorption. At 2,000 mg per liter of calcium, the suppression is approximately 19 percent. |
Quality Control Requirements
|
Calibrate instrument using calibration standards (CAL). Quality control samples (QCS) and laboratory blanks (LB) analyzed at a minimum of I each after every 10 samples. (Reference OFR 95-443). |
Sample Handling
|
Container Description: 500 mL Polypropylene bottle, wide-mouth. Treatment and Preservation: No treatment, preservation or special shipping requirements. |
Maximum Holding Time
|
180 days |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|
I-5485, P-0520 |