USGS-NWQL: I-5239: Cobalt, recoverable-from-bottom-material, various, FLAA
Official Method Name
|
Cobalt, atomic absorption spectrometric, direct |
---|---|
Current Revision
| Revised 1985 |
Media
|
VARIOUS |
Instrumentation
|
Flame Atomic Absorption |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
Methods for the Determination of Inorganic Substances in Water and Fluvial Sediments, Techniques of Water-Resources Investigations of the United States Geological Survey, Book 5, Chapter A1 Edited by Marvin J. Fishman and Linda C. Friedman |
Brief Method Summary
|
Cobalt is determined by atomic absorption spectrometry by direct aspiration of the sample into an air-acetylene flame without additional treatment of the sample other than the addition of ammonium chloride to mask certain interferences. |
Scope and Application
|
This method may be used to analyze water and water-suspended sediment containing at least 50 ug/L of cobalt. Samples solutions containing more than 1,000 ug/L need either to be diluted or to be read on a less expanded scale. Sample solutions containing less than 50 ug/L need to be analyzed by the atomic absorption spectrometric chelation-extraction method, providing the interference limits discussed in that method are not exceeded. This method may be used to analyze bottom material containing at least 5 ug/g of cobalt. Total recoverable cobalt in water-suspended sediment needs to undergo preliminary digestion-solubilization by method I-3485, and recoverable cobalt in bottom material needs to undergo preliminary digestion-solubilization by method I-5485 before being determined. |
Applicable Concentration Range
|
> 5 ug/g |
Interferences
|
Nitrate at 1 mg/L interferes by suppressing the absorption of the cobalt. This interference is eliminated in solutions containing about 18,000 mg/L of ammonium chloride. Samples adjusted to this concentration of ammonium chloride show no interference from 800 mg/L of nitrate. Individual concentrations of sodium (9,000 mg/L), potassium (9,000 mg/L), calcium (4,500 mg/L), magnesium (4,500 mg/L), sulfate (9,000 mg/L), chloride (15,000 mg/L), iron (4 X 106 ug/L), and cadmium, nickel, copper, zinc, lead, and chromium (10,000 ug/L) do not interfere. Higher concentrations of each constituent were not investigated. |
Quality Control Requirements
|
Calibrate instrument using calibration standards (CAL). Quality control samples (QCS) and laboratory blanks (LB) analyzed at a minimum of I each after every 10 samples. (Reference OFR 95-443). |
Sample Handling
|
Container Description: 500 mL Polypropylene bottle, wide-mouth. Treatment and Preservation: No treatment, preservation or special shipping requirements. |
Maximum Holding Time
|
180 days |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|
I-5485, P-0520 |