Modern Water: A00134:  PCB (polychlorinated biphenyl) in water by immunoassay

  • Summary
  • Analytes
  • Revision
  • Data and Sites
Official Method Name
PCB in Water RaPID Assay A00134 (SDI 1998) http://www.sdix.com/
Current Revision
1998
Media
WATER
Instrumentation
Immunoassay
Method Subcategory
Organic
Method Source
  Modern Water
Citation
  Modern Water Inc. User Guides
Brief Method Summary
Inhibition immunoassay, with a colorimetric indicator for competitive binding of analyte and an enzyme-labeled conjugate to surface-immobilized antibodies. Colorimetric response is inversely related to the (log) concentration of analyte in the sample or standard. Calibrated for Aroclor 1254. Reagents are purchased in kits with manufacturers' instructions for sample preparation and analysis. Requirements: a clean water supply to prepare dilutions of water samples; filtration equipment to remove particulates from water samples; an SDI magnetic bead separator, required for the SDI RaPID assay kit; a spectrometer which accepts sample tubes (SDI RaPID Assay) for analyses at visible wavelengths (450 nm primary; 600 or 650 nm reference); equipment to automate sample preparation for large numbers of samples (prefiltration, measurement of uL volumes, rinsing, agitation), so that contact times with reagents are controlled as required; computerized data acquisition and processing. Expendable supplies - approximately $8 per determination.
Scope and Application
Ambient and compliance monitoring. Can be adapted for analysis of soils by Method 4020. Screening for polychlorinated biphenyls by immunoassay (EPA 1995).
Applicable Concentration Range
0.5 - 10 ug/L as Aroclor 1254
Interferences
Sensitivity < Aroclor 1254: Aroclor 1242, Aroclor 1248, Aroclor 1260, Aroclor 1262. Sensitivity << Aroclor 1254:: Aroclor 1016; Aroclor 1221; Aroclor 1232; Aroclor 1268
Quality Control Requirements
Recommended for a kit allowing 100 quantitative determinations:

a. calibration with 3 standards and 1 blank, all analyzed in duplicate

b. precision: all 46 samples analyzed in duplicate (repeats)

c. accuracy: 3 matrix samples, spiked with the target analyte at different levels in the range for quantitative analysis.

d. validation: analysis of 4 positive and 4 negative samples by an independent method, for confirmation
Sample Handling
Samples are typically collected in glass containers with Teflon-lined caps. Immediate dilution of water samples with stabilizer (methanol:sample, 1:1, v/v) is recommended to inhibit adsorption of PCBs to the container. Drinking water samples are typically dechlorinated with 0.008% sodium thiosulfate, but the choice of reagents needs to be validated for the kit being used. Samples are held in a refrigerator at 4 oC. Immunoassay reagents and kits are stored in the dark at 2 to 8 oC until use.
Water samples and soil extracts generally need to be centrifuged or prefiltered to 0.2 um and diluted to 3 levels, in series. Aliquots of samples and reagents are accurately measured (mL) and mixed for required times, following manufacturer's instructions for use of the materials purchased.
Maximum Holding Time
5 days at 4oC until extraction and analysis
Relative Cost
Less than $50
Sample Preparation Methods