USGS-NWQL: I-6302:  Cyanide, recoverable-from-bottom-material, dry wt, colorimetric

  • Summary
  • Analytes
  • Revision
  • Data and Sites
Official Method Name
Cyanide, colorimetric, barbituric acid, automated-segmented flow
Current Revision
1985
Media
SOILS/SEDIMENT
Instrumentation
Automated Spectrophotometer
Method Subcategory
Inorganic
Method Source
  USGS-NWQL
Citation
Methods for the Determination of Inorganic Substances in Water and Fluvial Sediments, Techniques of Water-Resources Investigations of the United States Geological Survey, Book 5, Chapter A1 Edited by Marvin J. Fishman and Linda C. Friedman
Brief Method Summary
This method is based on the chlorination of cyanide with chloramine-T and on the subsequent reaction with a pyridine-barbituric acid reagent (Goulden and others, 1972). This method detects simple cyanides only; therefore, any complex cyanides must first be broken down by passing the acidified sample solution through an ultraviolet digestion-distillation procedure. The distillation step also removes certain interferences.
Scope and Application
This method may be used to determine cyanide in bottom material containing at least 0.5 mg/kg.
Applicable Concentration Range
at least 0.5 mg/kg
Interferences
Chloride interferes if its concentration exceeds 3,000 mg/L. Oxidizing agents may interfere. Glycine and urea at the 10-mg/L level do not interfere. A concentration of 10 mg/L sulfide increases the apparent cyanide concentration by approx 0.02 mg/L. Concentrations of sulfide greater than 10 mg/L interfere considerably. Thiocyanate is broken down to cyanide and sulfide by this procedure and, therefore, interferes on an equimolar basis. Sulfate concentrations of 4,000 mg/L do not interfere. Higher concentrations were not tested.
Quality Control Requirements
Calibrate instrument using calibration standards (CAL). Quality control samples (QCS) and laboratory blanks (LB) analyzed at a minimum of I each after every 10 samples
Sample Handling
Container Description: 500 mL Polypropylene bottle, wide-mouth Treatment and Preservation: No treatment, preservation or special shipping requirements
Maximum Holding Time
180 days
Relative Cost
Less than $50
Sample Preparation Methods