Standard Methods: 4500-F- E:  Fluoride by Complexone

  • Summary
  • Analytes
  • Revisions
  • Data and Sites
Official Method Name
4500-F- E. Complexone Method
Current Revision
Standard Methods Online
Media
WATER
Instrumentation
Automated Spectrophotometer
Method Subcategory
Inorganic
Method Source
  Standard Methods
Citation
  Standard Methods Online - Standard Methods for the Examination of Water and Wastewater
Brief Method Summary
The sample is distilled in the automated system, and the distillate is reacted with alizarin fluorine blue-lanthanum reagent to form a blue complex that is measured colorimetrically at 620 nm.
Scope and Application
This method is applicable to potable, surface, and saline waters as well as domestic and industrial wastewaters.
Applicable Concentration Range
0.1 to 2.0 mg-F-/L
Interferences
Interferences normally associated with the determination of fluoride are removed by distillation.
Quality Control Requirements
See Section 4020 Quality Control/Quality Assurance.
Sample Handling
Preferably use polyethylene bottles for collecting and storing samples for fluoride analysis. Glass bottles are satisfactory if previously they have not contained high-fluoride solutions. Always rinse bottle with a portion of sample.
Maximum Holding Time
28 days (See Section 1060)
Relative Cost
Less than $50
Sample Preparation Methods
Section 4500-F- B