USGS-NWQL: O-4024-03:  Volatile organic compounds, whole water, gas chromatography/mass spectrometry, heated purge and trap

  • Summary
  • Analytes
  • Revision
  • Data and Sites
Official Method Name
Volatile organic compounds, whole water, gas chromatography/mass spectrometry, heated purge and trap
Current Revision
2003
Media
WATER
Instrumentation
Gas Chromatography with Mass Spectrometry Detection
Method Subcategory
Organic
Method Source
  USGS-NWQL
Citation
Rose, D.L., and Sandstrom, M.W., 2003, Methods of Analysis of the U.S. Geological Survey National Water Quality Laboratory - Determination of Gasoline Oxygenates, Selected Degradates, and BTEX in Water by Heated Purge and Trap/Gas Chromatography/Mass Spectrometry, U.S. Geological Survey Water-Resources Investigations Report 03-4079, 39 p.
Brief Method Summary
Volatile organic compounds are purged from the sample matrix by simultaneously bubbling helium through a 25-mL aqueaous sample and heating at 65oC. The compounds are trapped in a tube containing suitable sorbent materials and then thermally desorbed into a capillary gas chromatographic column interfaced to a mass spectrometer system. Selected compounds are identified by using strict identification criteria, which include analyzing standard reference materials and comparing retention times and relative ratios of the mass spectra. Compounds are quantitated using interal standard procedures. Quantitation that is extrapolated less than the lowest calibration standard is qualified as "estimated" to signify the lower confidence in the extrapolated concentration. Compounds are not quantitated if they do not strictly adhere to identification criteria. Compounds identified with concentrations within the calibration range are reported without qualification, unless quality control or holding times are compromised.
Scope and Application
This method is suitable for the determination of gasoline oxygenates, selected degradates, and BTEX (benzene, toluene, ethylbenzene, and xylenes) at low concentrations in whole-water samples. The method is applicable to analytes that can be efficiently removed from the water matrix by heating and purging with helium.
Applicable Concentration Range
0.035 - 0.62
Interferences
Samples can be contaminated during collection or analysis. Several types of laboratory and field blanks are used in this method to identify sources of contamination. Possible contamination sources include vehicle exhaust fumes, industrial stack emissions, outgassing of solvents from vehicle carpets and upholstery, copier machines, paint, and cleaning solutions. Sampling equipment used at contaminated sites might contain residual contaminants if not cleaned properly. During sample preparation and analysis in the laboratory, samples can be contaminated by common extraction solvents that are present in the laboratory atmosphere.
Carryover contamination can occur when samples containing high concentrations of VOCs (greater than 20 ug/L) contaminate the next analysis at detectable concentrations because of residual VOCs in the trap, purge vessel, or transfer lines, which were not eliminated during the routine bake procedure. Samples suspected of being contaminated by carryover will be reanalyzed. The field-sampling personnel should note on the analytical service request form (ASR) if it is known that a given sample contains high concentrations of VOCs. The laboratory analysts should separate contaminated samples from clean samples. Knowledge of carryover characteristics by instrument and compound is necessary for this method.
Hydrogen sulfide will interfere with the response of the mass spectrometer and can damage columns, traps, multipliers, and quadropoles. If field personnel detects any odor of hydrogen sulfide, they should note this on the ASR.
Foamy samples can interfere with the analysis by raising the baseline, decreasing instrument response, and shifting peak retention times. Samples that are excessively foamy are diluted until no foam is produced.
Care needs to be taken to eliminate all potential organic contaminants from the volatiles laboratory by limiting exposure to laboratories using organic solvents, and exposed clothing.
Quality Control Requirements
Sample Handling
Samples known to contain bacteria adapted to degrading fuels should be preserved to pH 2 if analysis of BTEX compounds is desired and methyl acetate is not a concern.
Maximum Holding Time
14 days
Relative Cost
Sample Preparation Methods