USGS-NWQL: I-2522-90: Ammonia in Water by Colorimetry
Official Method Name
|
Nitrogen, Ammonia, Colorimetry, Salicylate-Hypochlorite, Automated-Segmented Flow |
---|---|
Current Revision
| 1993 |
Media
|
WATER |
Instrumentation
|
Spectroscopy (Colorimetry; Photometry) |
Method Subcategory
|
Inorganic |
Method Source
|
|
Citation
|
M.J. Fishman, 1993, Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory--Determination of inorganic and organic constituents in water and fluvial sediments: U.S. Geological Survey Open-File Report 93-125 |
Brief Method Summary
|
Ammonia reacts with salicylate and hypochlorite ions in the presence of ferricyanide ions to form the salicylic acid analog of indophenol blue (Reardon and others, 1966; Patton and Crouch, 1977; Harfmann and Crouch, 1989). The resulting color is directly proportional to the concentration of ammonia present. |
Scope and Application
|
This method is used to analyze samples of surface, domestic, and industrial water, and brines containing from 0.01 to 1.5 mg/L of ammonia-nitrogen. Concentrations greater than 1.50 mg/L must be diluted. This modified method was implemented in the National Water Quality Laboratory in March 1988. |
Applicable Concentration Range
|
0.01 - 1.5 mg/L (undiluted) |
Interferences
|
Sulfide interferes. Bromide and nitrite can interfere. Calcium and magnesium in highly alkaline waters (pH greater than 13.6) can exceed the ability of the tartrate to complex both ions.
The samples are easily contaminated by ammonia in the laboratory atmosphere; therefore, sample handling and analysis need to be performed where there is no possibility of ammonia contamination. |
Quality Control Requirements
|
Quality-control samples area analyzed at a minimum of one in every ten samples. These QC samples include at least one of each of the following: blanks, quality control samples, third party check solutions, replicates, and spikes. Correlation coefficients for calibration curves must be at least 0.99. QC samples must fall within 1.5 standard deviations of the mean value. If all of the data-acceptance criteria in the SOPs are met, then the analytical data are acceptable. |
Sample Handling
|
Description: 125 mL Brown polyethylene bottle. Treatment and Preservation: Filter through 0.45-um filter, use filtered sample to rinse containers, chill and maintain at 4oC, ship immediately. |
Maximum Holding Time
|
30 days from sampling |
Relative Cost
|
Less than $50 |
Sample Preparation Methods
|