Hach Co.: 8289:  Acid/Base Milliequivalent Determination Using Sulfuric Acid

  • Summary
  • Analytes
  • Revision
  • Data and Sites
Official Method Name
Acid/Base Milliequivalent Determination Using Sulfuric Acid
Current Revision
7/03
Media
WATER
Instrumentation
Titration
Method Subcategory
Inorganic
Method Source
  Hach Co.
Citation
  Hach Water Handbook (1997)
Brief Method Summary
A measured amount of sample is treated with a phenolphthalein colorimetric indicator and then titrated with a strong acid or base to pH 8.3. The amount of titrant used is directly proportional to the milliequivalents od acid or base in the sample. These titrations also cab be performed using a pH meter instead of a colorimetric indicator.
Scope and Application
For water, wastewater, and seawater with beginning pH greater than 8.3
Applicable Concentration Range
0-25000 meq/L
Interferences
Highly colored or turbid samples may mask the color change at the end point.
Quality Control Requirements
Sample Handling
Collect samples in clean plastic or glass bottles. Fill completely and cap tightly. Avoid agitation and prolonged exposure to air. Samples should be analyzed as soon as possible after collection but can be stored at least 24 hours by cooling to 4 deg. C or below. Warm to room temperature before analyzing.
Maximum Holding Time
24 hours @ 4deg. C.
Relative Cost
Less than $50
Sample Preparation Methods